Microbiology Devices; Reclassification of Human Immunodeficiency Virus Viral Load Monitoring Tests

Information

Type

Proposed Rule

Regulated products

Published

November 24, 2021

Last updated

November 24, 2021

Description

Document number 2021-25372. The Food and Drug Administration (FDA or the Agency) is proposing to reclassify human immunodeficiency virus (HIV) viral load monitoring tests, a postamendments class III device with the product code MZF, into class II (special controls), subject to premarket notification. FDA is also proposing a new device classification regulation along with special controls that the Agency believes are necessary to provide a reasonable assurance of safety and effectiveness for this device type. FDA is proposing this reclassification on its own initiative. If finalized, this order will reclassify this device type from class III (premarket approval) to class II (special controls) and reduce the regulatory burdens associated with these devices because manufacturers will no longer be required to submit a premarket approval application (PMA) for this device type but can instead submit a less burdensome premarket notification (510(k)) and receive clearance before marketing their device.