Anesthesiology Devices; Reclassification of Membrane Lung for Long-Term Pulmonary Support; Redesignation as Extracorporeal Circuit and Accessories for Long-Term Pulmonary/Cardiac Support

Information

Type

Proposed Rule

Regulated products

Published

January 8, 2013

Last updated

January 8, 2013

Description

Document number 2013-00086. On its own initiative, based on new information, the Food and Drug Administration (FDA) is proposing to reclassify membrane lung devices for long-term pulmonary support, a preamendments class III device, into class II (special controls) for conditions where imminent death is threatened by cardiopulmonary failure in neonates and infants or where cardiopulmonary failure results in the inability to separate from cardiopulmonary bypass following cardiac surgery. A membrane lung for long-term pulmonary support refers to the oxygenator component of an extracorporeal circuit used during long-term procedures, commonly referred to as extracorporeal membrane oxygenation (ECMO). Because circuit components used with the oxygenator are to be subject to the same regulatory controls, all of the device components used in an ECMO procedure are being considered in the scope of this proposed order, and the title and identification of the regulation will be revised accordingly to include extracorporeal circuit and accessories for long- term pulmonary/cardiac support.